Analytical method:
Winkler titration method using Langdon amperometric detection of the endpoint.
Samples preserved with 1 mL each of MnCl2 and NaI-NaOH solutions.
Shaken twice and stored in dark with tap water around caps to form seal.
Precipitate dissolved with 1mL H2SO4 solution and titrated with thiosulfate solution.
Endpoint is detected via electrode current that is related to concentration of I3-.
Normally run within 24 hours, but some samples from underway system sat for 2-3 days.
3-5 KIO3 standards run at beginning of each day running samples.
Blanks run in distilled water used for calculations.
Seawater blanks required 2-2.5 uL more titrant than distilled water blanks.
Calculations:
Final dataset calculated using distilled water blanks in keeping with official WOCE methodology.
High quality seawater blanks were determined on this cruise.
Data is likely of higher accuracy when calculated with seawater blanks.
To convert distilled blank numbers to seawater blank numbers, simply subtract 0.530 umol/kg (micromol/kg).
Saturations calculated relative to Garcia and Gordon (1992,1993) with no correction for atmospheric pressure variations.
BCO-DMO Processing Notes
- Generated from original file Winkler_SOGasEx_Underway.csv
BCO-DMO Edits
- date reformatted to YYYYMMDD
- Time reformatted to HHMM
- lon/lat padded to 3 decimal places
- Column headers/parameter names edited to conform to BCO-DMO convention (blanks replaced with underscores or eliminated)
- blank fields changed to 'nd'
- decimal places padded as appropriate for consistency
Comments from original text file - Winkler_SOGasEx_Underway.csv
Final Discrete Oxygen Data from underway system analyzed by Winkler Titration. SOGasEx cruise. Feb 29-April 11, 2008 aboard R/V Ron Brown.
This data is calculated with a distilled water blank. To obtain values calculated with a seawater blank, subtract 0.530 umol/kg.
Flag = 2 means good; Flag = 3 means questionable (unexplained high std dev); Flag = 4 means bad (see metadata file for explanation).