<pre>
Methods:
In general, the methods employed for the bottle salinity, Winkler
dissolved oxygen, and nutrient analyses did not differ significantly
from those described in the JGOFS protocols that were distributed in
June, 1994. Minor differences included the following: 1) Sea Bird CTD
systems and bottle carousels were employed (SBE- 9+ underwater units,
SBE-11 deck units, SBE-32 carousels). These units represent a newer
generation of equipment than the units described in the JGOFS
protocols. 2) The weights of the salts used for primary standards for
dissolved oxygen and nutrients were not adjusted to an "in vacuo"
basis as suggested in the protocols. It is unlikely that this
departure from procedure would cause significant errors. Our
calculations suggest that the maximum differences arising from our
decision to not correct to an "in vacuo" basis would range from 0.02%
(oxygen standards) to 0.06%(ammonium standards). 3) The protocols
give one a choice of adjusting nutrient methods so that calibration
curves are strictly linear, or opting for more response and taking
into account non-linearities. We choose the latter method. 4) No
corrections were made for "carryover" between nutrient samples run on
the Technicon Autoanalyzer. Data from this cruise, suggest that
carryover effects in our nutrient analyses are generally less than ~2%
of the concentration difference between adjacent samples. Examination
of cases where more than one sample was taken from a depth at which
there was a significant increase in nutrient concentrations will help
the user determine the carryover effect for many individual casts. 5)
Calibration and re-calibration of volumetric ware were not exactly as
described in the JGOFS protocols, but this was largely compensated for
by comparing independent standards diluted with independent volumetric
ware, and by re-calibration of some of the volumetric ware after
cruises TN045 and TN050. DATA FOR THIS CRUISE (TN049) HAVE BEEN
CORRECTED FOR ERRORS IN THE PIPETS BY MULTIPLYING THE SILICATE VALUES
BY 0.999. 6) Duplicate oxygen samples were not drawn from every Niskin
or Go-Flo bottle, but there were several comparisons of bottles
tripped at the same depth. 7) Azide was added to the Winkler oxygen
pickling reagents to destroy nitrite that can be present in relatively
high concentrations in the Arabian Sea.
Temperature:
The temperature data associated with each bottle depth were taken by
the CTD system during the bottle tripping process. Consult the
companion CTD data report for this cruise to learn more about the CTD
system.
Sampling:
The samples in this report were taken from 10 liter Niskin
bottles.
Because there is little or no lag time between triggering a bottle and
bottle closure with the new SeaBird rosette systems, bottles were
generally held at the sampling depth for at least 30 seconds before
tripping or until the deck read-outs stabilized if this took more than
30 seconds.
NOTE THAT THE MID-POINT OF THE SAMPLING BOTTLES WAS ONE METER ABOVE
THE CTD SENSORS. THE DATA HAVE NOT BEEN CORRECTED FOR THIS ONE METER
OR 1.1 DECIBAR DIFFERENCE BETWEEN CTD SENSOR AND SAMPLING BOTTLE
POSITIONS.
Salinity:
Salinities were determined with Guildline Autosal salinometers. New
vials of standard sea-water were used to standardize before and at the
end of every run (12-36 samples). Agreement between bottle salinities
and the recently calibrated sensors on the Sea Bird CTD systems was
usually better than 0.01 (except in regions of strong gradients)
before post-cruise data processing which employs the bottle salinities
to correct the CTD salinities. More information on the quality of the
salinity data are given in the companion CTD report. Both the CTD
salinity data at the time of bottle tripping and the salinities run on
the Niskin bottle samples with an Autosal salinometer are reported
here.
Dissolved oxygen:
The Winkler dissolved oxygen set-up was built and supplied by the
SIO/ODF group. This system is computer controlled and detects the
end-point photometrically. Temperature of the thiosulfate and
standard solutions is automatically monitored by this system. Checks
on cruises TN039 and TN043 between independent standards prepared with
independent volumetric ware gave agreement of +-0.02 per cent. A
similar check made during TN054 suggested agreement of ~+-0.15 per
cent. The linearity of the "Dosimat" automatic buret was also checked
during cruises TN043 and TN054 with good results.
Nutrients:
Note that the terminology used to describe nutrients has become
somewhat loose over the years and that silicate=silicic acid, and
phospate=reactive phosphorus.
Nutrient analyses were performed on a 5-channel Technicon II AA
system that was modified and provided by the SIO/ODF group.
In assessing the nutrient standard comparisons outlined below,
note that the full-scale ranges for nutrients were as follows:
Ammonium =0 to 5 micromolar
Nitrate =0 to 45 "
Nitrite =0 to 5 "
Phosphate =0 to 3.6"
Silicate =0 to 180 "
These ranges were arrived at after an Internet pole of PI's and
cover the full depth concentration range for the Arabian Sea.
On the set-up and calibration cruise (TN039), the SIO/ODF nitrate and
nitrite standards and standards from the National Institute of
Oceanography in India (provided by S.W.A. Naqvi) were compared with
the following results:
NIO Nitrate Std.= 22.6 micromolar; SIO/ODF=22.5 micromolar
NIO Nitrite Std.= 2.42 micromolar; SIO/ODF = 2.50 micromolar
As can be inferred from the above, the nitrate plus nitrite
values were almost identical in the mixed standards;
25.02 (NIO) vs 25.00 (SIO)micromolar.
On TN039 Lou Codispoti prepared independent primary nitrate, nitrite,
silicate and phosphate standards for comparison with SIO/ODF primary
standards, and made dilutions using glassware entirely independent of
the SIO/ODF glassware.
The results were as follows:
Codispoti SIO/ODF
Nitrate 26.96 micromolar 26.85 micromolar
Nitrite 2.90 " 2.86 "
Silicate 86.4 " 85.8 "
Phosphate 2.36 " 2.36 "
All of the above results are within plus or minus 0.5% of the full
scale values, and with the exception of nitrite, the rest are within
plus or minus 0.2% of the full scale values. On TN043, the volumetric
equipment used for making routine nitrate and phosphate standards was
checked against volumetric ware calibrated by LAC. The average of the
results agreed to within +-0.1% of the full scale value for phosphate
and +-0.2% of the full scale value for nitrate.
Because nitrite values in the suboxic waters of the Arabian Sea can
attain values of approximately 5 micromolar, we kept track of the
efficiency of the Cd column that reduces nitrate to nitrite in the
nitrate analysis. The efficiencies were all greater than 99% during
this cruise, so no corrections have been made for any errors in
nitrate arising from deviations in cadmium column efficiency. NOTE
THAT THE FULL-SCALE NITRITE RANGE FOR THIS CRUISE WAS 5 MICROMOLAR AND
THAT SOME CONCENTRATIONS EXCEEDED THIS VALUE. IN THESE CASES, THE
SAMPLES EITHER HAD TO BE DILUTED OR THE VOLTAGE RANGE CHANGED ON THE
RECORDER. THESE MANIPULATIONS TEND TO DEGRADE THE ACCURACY OF NITRITE
VALUES IN EXCESS OF 4.5 TO 5 MICROMOLAR.
The ammonium results are the least precise of all the nutrient
results. On TN039, three primary standards were compared with
agreement of about plus or minus three per cent of the full-scale
value. These standards may have agreed within the precision of the
method, but we found a significant salinity effect on the ammonium
results that might explain some of these differences since the
salinities of the comparison standards varied a bit. Experiments on
this first JGOFS Arabian Sea process study cruise (TN043) suggest that
the ammonium signal decreases by approximately 3.5% for a salinity
increase of 1.00. Comparisons of an independent standard compared by
LAC with the SIO standard on this cruise (TN043) when corrected for
salinity differences between the standards agreed to ~ + -0.1% of the
full-scale value. The largest absolute difference was 0.025
micromolar and the average difference was 0.013 micromolar for six
comparisons between 1-3 micromolar. Thus, the average difference
between these two independent standards was + -0.006 micromolar. These
results tend to confirm the need to take salinity differences between
samples and standards into account when calculating the final ammonium
concentrations. THE AMMONIUM VALUES IN THIS REPORT HAVE BEEN CORRECTED
FOR THIS EFFECT. ON THIS CRUISE THE SALINITY EFFECT CORRECTION IS A
2.7% DECREASE IN SIGNAL FOR A SALINITY INCREASE OF 1.00. On this
cruise, the salinity of the working standards used to calibrate the
ammonium method was ~35.14.
</pre>